Introduction
JC30 is an industrial measuring instrument that is modular in design and meets global OEM standards. The series is stable and reliable, and has excellent cost performance. It is rugged and durable to meet the requirements of more stringent industrial standards. It is widely used in industrial support, industry support and equipment support. The series includes a variety of options, which are applied to different industries and different working conditions. It solves the needs for economic pressure monitoring in different occasions and serves a wide range of industries.
JC30 air pressure / micro differential pressure transmitter adopts new technology to produce micro range, high stability silicon piezoresistive non-oil-filled sensor components and special digital integrated circuits. It is firmly packaged with 304 stainless steel casing through stress isolation technology. The measured air pressure or differential pressure enters the positive and negative pressure chamber of the sensor through the G1/4 external thread or φ6 integrated barbed tower pressure interface, and the corresponding air pressure or differential pressure is converted into a plurality of industrial standard signal outputs proportional to the sensor detection, Such as 4 ~ 20mA, 0-5V DC, 0.5-4.5V DC.
Features
* Measuring range: 0~100Pa...100KPa
* All stainless steel construction, compact and durable, easy to install
* Temperature range is -20~100°C
* Wind pressure, differential pressure two kinds of structural interoperability design
* Accuracy 1-2% FS for pressure below 250Pa, accuracy 0.5% of industry standard FS for other ranges
* Reliable quality and cost-effective
Application
* HVAC
* Energy Management System
* VAV and fan control
* Environmental pollution control
* Clean room, clean room project
* Oven pressurization and furnace ventilation control
* Monitoring of natural gas and gas pipeline network
* Downhole ventilation test
* Pharmaceutical equipment and medical equipment
* Filter pressure
* Building exhaust system
Performance Parameter
Measuring Range |
0~100Pa...100KPa |
Overload Capability |
1.5~2 times full-scale pressure |
Burst Pressure |
4xFS |
Durability |
﹥1x107 cycle(P:0~FS) |
Pressure Type |
Gauge / Absolute |
Measuring medium |
Dry clean gas compatible with 304 stainless steel |
Response time |
≤4ms |
Resolution |
Minimum. 0.5Pa |
Accuracy (linear,hysteresis, repeatability) |
Typical:±0.5%FS |
Maximum:±1%FS |
Typical:±1%FS(below 250Pa) |
Maximum:±2%FS(below 250Pa) |
Long-term stability |
Typical:±0.2%FS/year |
Maximum:±0.3%FS/year |
Zero temperature drift |
Typical:±0.02%FS/℃ |
Maximum:±0.05%FS/℃ |
Sensitivity temperature drift |
Typical:±0.02%FS/℃ |
Maximum:±0.05%FS/℃ |
Environmental Conditions
Medium Temperature |
-40~100℃ |
Ambient Temperature |
-40~85℃ |
Compensation Temperature |
-20~80℃ |
Vibration resistance |
10g IEC 60068-2-6 |
Impact resistance |
100g/1ms IEC 60068-2-27 |
EMC- launch |
EN61000-6-3 |
EMC- anti-interference |
EN61000-6-2 |
Insulation resistance |
﹥100MΩ/100V |
Shell Protection |
Plug type(IP65); Cable type(IP67) Compliance IEC 60529 standard |
Certification |
CE |
Electrical Specifications
Code |
Standard signal(with
short circuit protection) |
Supply voltage with
polarity protection |
Power supply-Current |
Load(R) |
Output Impendence |
A1 |
4~20mA |
DC 9~30V |
Max.25mA |
R≤(U-9)/0.02Ω |
|
V1 |
1~5V DC |
DC 9~30V |
8mA |
R≥50kΩ |
<2kΩ |
V2 |
0~5 V DC |
DC 9~30V |
8mA |
R≥50kΩ |
<2kΩ |
V3 |
0.5~4.5V DC |
DC 9~30V |
8mA |
R≥50kΩ |
<2kΩ |
V4 |
0.5~4.5V DC |
DC 5±0.25V |
8mA |
R≥50kΩ |
<2kΩ |
Material
Interface and housing |
Stainless Steel 304L |
O-ring |
Silicone Rubber |
Sensor Diaphragm |
Silicon wafer(measured dry ,dust-free, non-corrosive or weakly corrosive, clean gas) |
Weight |
Approx.150g |
JC30 |
G |
01KN |
A1 |
F2 |
W2 |
Model |
Pressure type |
Pressure Range |
Output |
Mounting thread |
Electrical connections |
JC30 |
G=Gauge or wind/
air pressure
D=Differential Pressure |
002C= ±200Pa G/D
005C= ±500Pa G/D
01KC= ±1KPa G/D
02KC= ±2KPa G/D
05KC= ±5KPa G/D
10KC= ±10KPa G/D
005N= -500Pa G/D
01KN= -1KPa G/D
02KN= -2KPa G/D
05KN= -5KPa G/D
10KN= -10KPa G/D
20KN= -20KPa G/D
002P= 200Pa G/D
005P= 500Pa G/D
01KP=1KPa G/D
02KP= 2KPa G/D
05KP= 5KPa G/D
10KP= 10KPa G/D
20KP= 20KPa G/D
40KP= 40KPa G/D
001B= 100KPa G/D |
A1=4~20mA
V1=1~5V
V2=0~5V
V3=0.5~4.5V
V4=0.5~4.5V/5V |
F1=M20x1.5male
F2=G1/4male
F5=Ф8 gas nozzle
F0= Customize |
W1=Straight Out 1m
W2=A type HSM plug
W3=M12 corner out 1m
W4=M12 Straight Out 1m |
Model example:
JC30G01KNA1F2W3
(JC30 Air Pressure/ Micro Differential Pressure Transmitter; Range 0~1MPa Gauge; Output 4~20mA; Accuracy 0.5% typical; Power Supply 9~30VDC;Pressure Connection G1/4 male thread; Electrical Connection M12 corner out 1m; Selection Remarks: The range meter 01KP=1KPa G/D G indicates the gauge pressure or wind/air pressure, which is the pressure difference with the atmosphere, that is, the wind pressure measurement of a single pressure port, D is the differential pressure measurement, measuring the pressure difference between the two pressure ports. Use the B-copper joint in the outline drawing to connect the negative pressure port, as shown in the outline drawing.
Remarks: It can be customized if the order quantity exceeds a certain amount. Please contact the sales engineers for details.
Product reviews can be carried out after the member login, please